BE multi-biosensor module for the quantitative measurement of sodium, potassium, and chloride in serum, plasma, and urine, designed for electrolyte analyzers.

B&E

Manufacturer

Description

The CBS-3/30/300 multi-biosensor module provides quantitative measurement of Na⁺, K⁺, and Cl⁻ ions in serum, plasma, urine (diluted or undiluted), cerebrospinal fluid, pleural effusion, and ascitic fluid. Its electrochemical slope (40 to 70 mV under normal use) and precision (CV ≤ 1.5%) ensure reliable results across a wide linearity range. For in vitro diagnostic use only, the module must rest for 1 hour at room temperature after removal from cold storage before installation. Stored between 2 and 30 °C, it remains valid for 1 year (1 month after opening the packaging) and provides service for at least 600 samples over 20 days of use. Whole blood is not compatible with this module (risk of precipitates clogging the fluid circuits).